Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group
F Udink ten Cate1, J Breur, N Boramanand
1Wilhelmina Children's Hospital, Utrecht, Netherlands.
Insights
Steroid eluting endocardial leads showed better stimulation thresholds in pediatric patients requiring permanent pacing compared to epicardial leads. However, both lead types demonstrated comparable overall performance and longevity in this population.
Area of Science:
- Pediatric Cardiology
- Cardiac Electrophysiology
- Biomedical Engineering
Background:
- Permanent pacing in infants and children is crucial for managing bradyarrhythmias.
- Steroid-eluting leads are designed to reduce pacing threshold rise and improve longevity.
- Comparing epicardial and endocardial lead performance in pediatric patients is essential for optimizing device selection.
Purpose of the Study:
- To compare the performance of steroid-eluting epicardial versus endocardial leads in pediatric patients.
- To evaluate pacing and sensing characteristics, impedances, and lead longevity.
- To assess complication rates associated with each lead type.
Main Methods:
- A retrospective evaluation of 159 steroid-eluting leads implanted in 95 children.
- Patients were categorized into two groups based on weight: Group A (<15 kg) and Group B (>15 kg).
- Pacing thresholds, sensing values, impedances, and complication rates were analyzed for both endocardial and epicardial leads.
Main Results:
- Ventricular endocardial leads demonstrated significantly lower stimulation thresholds at implant and follow-up in both weight groups (p < 0.014 and p < 0.001, respectively).
- Sensing thresholds were also better for ventricular endocardial leads at follow-up in Group B (p < 0.02).
- Complications requiring intervention were observed in both groups, with a higher incidence in the epicardial lead group (n=18) compared to the endocardial group (n=7).
Conclusions:
- Steroid-eluting endocardial leads offer superior pacing thresholds in pediatric patients compared to epicardial leads.
- Despite differences in stimulation thresholds, both endocardial and epicardial steroid-eluting leads exhibit comparable overall performance and longevity in the pediatric population.
- Lead choice should consider patient weight and specific pacing needs, balancing performance with complication risks.
Aim:
To compare the performance of steroid eluting epicardial and endocardial leads in infants and children requiring permanent pacing.
Methods:
Evaluation of pacing and sensing characteristics, impedances, and longevity of 159 steroid eluting leads implanted in 95 children. Group A consisted of 24 children weighing less than 15 kg with 15 endocardial leads (five atrial, 10 ventricular) and 19 epicardial leads (five atrial, 14 ventricular). Group B consisted of 71 children weighing more than 15 kg with 106 endocardial leads (56 atrial, 58 ventricular) and 19 epicardial leads (nine atrial, 10 ventricular).
Results:
Group A: Stimulation thresholds were lower for ventricular endocardial leads at implant (mean (SD) 0.84 (0.54) v 1.59 (0.64) V, p < 0.014) and at two year follow up (ventricular 0.64 (0.24) v 1.65 (0.69) V, p < 0.003). Impedance and sensing thresholds did not differ significantly at implant and follow up. Group B: Stimulation thresholds were lower for ventricular endocardial leads at implant (0.72 (0.48) v 1.48 (0.58) V, p < 0.001) and at follow up (0.88 (0.46) v 1.55 (0.96) V, p < 0.009). Impedance did not differ. Sensing thresholds were also better for ventricular endocardial leads at follow up (9.1 (5.2) v 14.2 (6.4) mV, p < 0.02). Complications requiring intervention occurred in both groups (n = 7 for endocardial v n = 18 for epicardial leads).
Conclusions:
Endocardial and epicardial steroid eluting leads have comparable performance in the paediatric population.