[Surgical and interventional use of radiofrequency current: is there interference with implantable
M Fiek1, U Dorwarth, I Durchlaub
1Medizinische Klinik I, Klinikum Grosshadern Ludwig-Maximilians-Universität München Marchioninistr. 15 81377 München, Germany.
Background:
During surgical and interventional procedures, interactions between implantable cardioverter defibrillators (ICD) and electrical cautery, respectively, application of radiofrequency (RF) energy may occur. Induction of inadequate shock therapies or device malfunction may result and represent a potential perioperative hazard for the patient.
Methods:
Hence, we analyzed the intraoperative interactions in 23 consecutive ICD patients with regard to different surgical and interventional procedures. Sixteen surgical operations (general surgery n = 7, urologic n = 5, abdominal n = 2, gynecological n = 1, thoracic n = 1) and 7 interventional therapies (RF catheter ablation n = 5, endoscopic papillotomy n = 2) were performed. The ICD devices were all located in the left pectoral position and consisted of 15 single and 8 dual chamber defibrillators. During the procedure tachyarrhythmia detection (VF 295 +/- 21 ms, VT 370 +/- 55 ms) of the devices was maintained active (monitoring mode); only ICD therapies were inactivated. The indifferent electrode of the electrical cauter/RF generator was placed in standard positions (right mid femoral position n = 18, thoracic spine area n = 5). After the procedure, the ICD memory was checked for detections, respectively, for changes of the programming.
Result:
There was no misdetection or reprogramming of the ICD caused by electrical cautery or RF energy.
Conclusions:
Despite the lack of undesired interactions ICDs should be inactivated preoperatively to assure maximum patient safety. However, should inactivation be ineffective or not manageable, electromagnetic interference is highly unlikely.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Dysrhythmias VI: Management of Dysrhythmias