Related Experiment Video
Updated: Aug 26, 2026

A Recovery Cardiopulmonary Bypass Model Without Transfusion or Inotropic Agents in Rats
Published on: March 23, 2018
Electrical failure during cardiopulmonary bypass: an evaluation of incidence, causes, management and guidelines for
M Hargrove1, B C Ramish, A O'Donnell
1Cardiothoracic Surgery Unit, Cork University Hospital, Wilton, Ireland. hargrovem@shb.ie
Abstract:
The incidence of electrical failure during cardiopulmonary bypass (CPB) has been reported to occur in approximately 1 per 1000 cases. While the resultant morbidity and mortality is low, electrical failure is a life-threatening scenario. We report three major electrical failures during CPB in a patient population of 3500 over a 15-year period. These cases involved mains failure and generator shut down, mains failure and generator power surge, and failure of the uninterruptable power supply (UPS), which caused protected sockets to shut down. Protocols for preventative maintenance, necessary equipment, battery backup and guidelines for the successful management of such accidents during CPB are discussed.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Aneurysm IV: Nursing Management