Related Experiment Videos
Technique for retrieval of a guidewire lodged in a vena cava filter
Michael Wholey1, Boulos Toursarkissian, George Velez
1Department of Radiology, University of Texas Health Science Center at San Antonio, TX 78229, USA.
Vascular and Endovascular Surgery
|September 24, 2002
Abstract:
A case in which the guidewire used to place a percutaneous jugular central line became entangled in the limbs of a Trapease inferior vena cava filter is presented. A 7F sheath introducer was fashioned into a monorail system to uncurl the wire's J tip and remove the wire without disturbing the filter's position. This report highlights an increasingly frequent clinical scenario and an inexpensive effective method for resolution.