Influence of reagents formulation on real-time PCR parameters
Related Experiment Video
Updated: Sep 29, 2026

Polymerase Chain Reaction: Basic Protocol Plus Troubleshooting and Optimization Strategies
Published on: May 22, 2012
J S Burgos1, C Ramírez, R Tenorio
1Centro de Biología Molecular Severo Ochoa - CSIC Universidad Autónoma de Madrid, Cantoblanco, Madrid, Spain. jburgos@cbm.uam.es
Abstract:
Real-time polymerase chain reaction (PCR) techniques are increasingly used to quantify target sequences for diagnostic and research purposes. Due to its 'quantitative' character, it is very important to determine the variability of this technique correlating with several experimental conditions. The objective of this study was to analyse the effect of manufacturing lots of PCR reagents on two main PCR parameters, specificity and sensitivity. For this study, we used four different amplicons, using either mouse genomic DNA or viral DNA. Although a PCR product could be obtained in any of the conditions, we observed that there are relevant variations in sensitivity depending on the reagents formulation. We conclude that different lots of reagents may determine the analytical performance of PCR assays indicating that reagents testing are of special importance when the PCR protocol is used for quantitative purposes.