Vertical acetabular positioning with an inclinometer in total hip arthroplasty
Related Experiment Video
Updated: Sep 29, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Pascal-André Vendittoli1, Nicolas Duval, David James Stitson
1Department of Orthopaedic Surgery, Centre Hospitalier Maisonneuve-Rosemont (CHA), Montréal, Québec, Canada. pa_vendittoli@hotmail.com
Abstract:
Vertical acetabular implant positioning is an important technical aspect in total hip arthroplasty. To evaluate the potential benefit of an inclinometer, 50 cup insertions were done on a cadaver pelvis. Acetabular cup vertical angles averaged 44.4 degrees +/- 11.4 degrees by visuospatial perception and 42.2 degrees +/- 3.8 degrees with the inclinometer. All cups were within the safe angle range of 40 degrees to 49 degrees with the inclinometer compared with 64% of cups by visuospatial perception. Use of the inclinometer reduced variability by a factor of 2.0 to 4.5. The addition of an inclinometer for acetabular cup insertion increases the probability of positioning the cup within a vertical safe range during total hip arthroplasty on a cadaver pelvis, suggesting that it could be a useful adjunct in clinical practice.