Quantification of fetomaternal hemorrhage by fluorescence microscopy is equivalent to flow cytometry
Nadin Ochsenbein-Imhof1, A F Ochsenbein, B Seifert
1Department of Obstetrics, University Hospital Zurich, Switzerland.
Background:
The quantification of fetal cells in the maternal circulation remains an important goal to determine the amount of anti-D necessary to prevent active immunization of a D- mother giving birth to a D+ baby. Underestimation of fetomaternal hemorrhage (FMH) results in inefficient anti-D prophylaxis and maternal immunization; overestimation of FMH results in higher doses of passively transferred anti-D, higher costs, and the risk of disease transmission. Thus, a reliable method to quantitatively assess FMH is necessary.
Study Design And Methods:
Serial dilutions of artificial FMH were quantitatively measured by three different methods: flow cytometry, fluorescence microscopy (each after anti-D staining), and by the Kleihauer-Betke test. The accuracy and precision of the three methods were compared by statistical analysis.
Results:
Fluorescence microscopy and flow cytometry were comparably accurate and precise in quantifying FMH. In contrast, the accuracy of the Kleihauer-Betke test was poor, resulting in substantial overestimation of FMH in the samples with lower fetal cell concentrations.
Conclusion:
Anti-D flow cytometry and fluorescence microscopy for detection of fetal cells offer equally reliable and precise methods in contrast to the Kleihauer-Betke test. Fluorescence microscopy may be established as standard to quantify FMH in clinical practice because it is comparable to flow cytometry; in addition, it is time saving and is less expensive.