Related Experiment Video
Updated: Mar 24, 2026

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Efficiency of a radiophotoluminescence glass dosemeter for low-earth-orbit space radiation
1National Institute of Radiological Sciences, Chiba-shi, Japan. h_yasuda@nirs.go.jp
Abstract:
Chips of a radiophotoluminescence glass dosemeter (RPLG) were used for measurements of space radiation during a 9.8 d Shuttle-Mir mission (STS-91) at an altitude of 400 km and an inclination of 51.65 degrees. Two of RPLG chips were put into each of 59 positions in or on a life-size human phantom. The RPLG values equivalent to 137Cs gamma ray absorbed doses were found to be systematically lower than those of a Mg2SiO4:Tb thermoluminescence dosemeter (TDMS). In comparison with the organ or tissue absorbed dose and dose equivalent values that were estimated using a combination of TDMS and plastic nuclear track detectors, the efficiencies of the RPLG chips were about 80% for the water absorbed dose and about 40% for the dose equivalent. Whereas the percentage values will change during different missions, such additional information obtained from small RPLG chips is useful for improving the reliability of radiation dosimetry in space.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:18Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
Published on: December 14, 2017
Related Concept Videos
Determination of Crystal Structures
Photoluminescence: Applications
Biological Effects of Radiation
Absorption of Radiation