Related Experiment Video
Updated: Sep 28, 2026

Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
Pitfalls in cefepime titration from human plasma: plasma- and temperature-related drug degradation in vitro
Denis Bugnon1, Eric Giannoni, Paul Majcherczyk
1Division of Infectious Diseases, Department of Internal Medicine, Centre Hospitalier Universitaire Vaudois, 1011 Lausanne, Switzerland.
Abstract:
While developing a high-pressure liquid chromatography assay for cefepime in plasma, we observed significant drug degradation at 20 and 37 degrees C but not at 4 degrees C. This plasma-related degradation persisted after protein removal. This warrants caution regarding cefepime assays for pharmacokinetic and pharmacodynamic studies of cefepime in vitro and in vivo.