Biomechanical properties of high-density polyethylene for pterional prosthesis
Related Experiment Video
Updated: Sep 28, 2026

Generation of Greater Bacterial Biofilm Biomass using PCR-Plate Deep Well Microplate Devices
Published on: April 22, 2022
Hun K Park1, Manuel Dujovny, Fernando G Diaz
1Department of Neurosurgery, School of Medicine, Wayne State University, School of Medicine, Detroit, MI 48201, USA. hpark@med.wayne.edu
Abstract:
The pterional approach is the most popular surgical technique in aneurysm and skull base tumor removal. Reconstruction of the temporal contour deformity due to craniotomy requires graft implantation. Porous high-density polyethylene (PHDPE) as a craniofacial and pterional implant material recently became available. However, material properties of the pterional implant are not yet known. In order to measure the biomechanical properties of PHDPE, we implemented the tensile test, the three-point bending test and the water displacement method for density measurement. Elastic modulus varies from 227 to 307MPa. Density range is 0.68 and 0.7 depending on the size of pores. The data can be used to study the character of the porous high-density polyethylene implant, how it resists stress or fatigue in combination with conventional plating systems.