Primary standardization of assays for anticonvulsant drugs: comparison of accuracy and precision
John F Wilson1, Ian D Watson, John Williams
1Department of Pharmacology, Therapeutics & Toxicology, University of Wales College of Medicine, Cardiff CF14 4XN, United Kingdom. wilsonjf@cardiff.ac.uk
Background:
The accuracy and precision of methods for the measurement of the anticonvulsants phenytoin, phenobarbital, primidone, carbamazepine, ethosuximide, and valproate in human serum were assessed in 297 laboratories that were participants in the United Kingdom National External Quality Assessment Scheme (UKNEQAS).
Methods:
We distributed lyophilized, serum-based materials containing low, medium, and high weighed-in concentrations of the drugs. The 297 participating laboratories received the materials on two occasions, 7 months apart. Expected concentrations were determined by gas chromatography or HPLC methods in five laboratories using serum-based NIST reference materials as calibrators.
Results:
In general, bias was consistent across concentrations for a method but often differed in magnitude for different drugs. Bias ranged from -1.9% to 8.6% for phenytoin, -2.7% to 3.1% for phenobarbital, -2.7% to 0.5% for primidone, -8.6% to 0.3% for carbamazepine, -5.6% to 2.0% for ethosuximide, and -7.2% to 0.1% for valproate. Intralaboratory sources of imprecision significantly exceeded interlaboratory sources for many drug/method combinations. The mean CVs for intra- and interlaboratory errors for the different drugs were 6.3-7.8% and 3.3-4.2%, respectively.
Conclusions:
For these long-established and relatively high-concentration analytes, the closed analytical platforms generally performed no better than open systems or chromatography, where use of calibrators prepared in house predominated. To improve the accuracy of measurements, work is required principally by the manufacturers of immunoassays to ensure minimal calibration error and to eliminate batch-to-batch variability of reagents. Individual laboratories should concentrate on minimizing dispensing errors.
More Related Videos
07:52Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
08:24Development and Standardization of an Ex Vivo Micromethod for Intracellular Quantification of Vincristine in Primary ALL Cells by LC-MS/MS
Published on: January 23, 2026
Related Concept Videos
Drug Dissolution: Requirements and Profile Comparison
Drug Concentrations: Measurements
Plasma —...
Bioequivalence Data: Statistical Interpretation
Therapeutic Drug Monitoring: Drug Analysis Methods
Drug Product Performance: In Vitro–In Vivo Correlation
In Vitro Drug Dissolution: Compendial Testing Models I