Related Experiment Video
Updated: Jul 31, 2026

Luminescence Lifetime Imaging of O2 with a Frequency-Domain-Based Camera System
Published on: December 16, 2019
Effect of 176Lu background on singles transmission for LSO-based PET cameras
J S Huber1, W W Moses, W F Jones
1Lawrence Berkeley National Laboratory, University of California, Berkeley 94720, USA.
Abstract:
We explore how the radioactive background from naturally occurring 176Lu affects single photon transmission imaging for lutetium orthosilicate (LSO) scintillator-based PET cameras by estimating the transmission noise equivalent count rate (NECR) including this background. Assuming a typical PET camera geometry (80 cm detector ring diameter), we use a combination of measurement and analytic computation to estimate the counting rates due to transmission, scatter and background events as a function of singles transmission source strength. We then compute a NECR for singles transmission. We find that the presence of radiation from the naturally occurring 176Lu reduces the NECR by 60% or higher for source strengths less than 10 mCi, and that a 25% reduction of the NECR can occur even with a source strength of 40 mCi.