Improvement of external fixator performance in type C pelvic ring injuries by plating of the pubic symphysis: an
K J Ponson1, G A Hoek van Dijke, Pieter Joosse
1Department of Traumatology, Academisch Medisch Centrum, Amsterdam, The Netherlands. k.j.ponson@amc.uva.nl
Background:
In an earlier study, we introduced a pelvic ring stability criterion for weightbearing stabilization. In a loading test, however, current external fixation systems alone did not meet this criterion. Internal fixation of the dorsal ring can significantly increase stability, but the condition of severely injured patients is often a contraindication for major surgery. The aim of this study is to optimize external pelvic ring fixation without dorsal ring stabilization to allow weightbearing in early mobilization of patients with unstable pelvic ring injuries.
Methods:
The stiffness of external fixation systems alone and in combination with one or two anterior plates was measured by using a pelvic replica with a type C pelvic ring injury. Endpoints were 15 mm of dislocation or tolerance of 560 N.
Results:
Addition of one plate at least doubles stiffness, whereas two-plate fixation results in at least a fourfold stiffer configuration. Frame configurations profit more than single-bar systems, and all but one system resist the weightbearing load after double-plating of the pubic symphysis.
Conclusion:
The choice of double-plate fixation of the anterior ring in addition to external fixation results in weightbearing capacity.
More Related Videos
Related Concept Videos
General External Flow Characteristics
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Internal and External Forces
External and Internal Respiration
External Anatomy of the Kidney
The kidneys are located in the retroperitoneal space on either side of the vertebral column, protected posteriorly by the 11th and 12th ribs. The right kidney sits slightly lower than the left owing to the presence of the liver...