[Flow through an improved ball valve for the heart assist device]
Related Experiment Video
Updated: Sep 28, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
D Aguilera1, K Affeld, Ch Lederer
1Universitätsklinikum Charité, Labor für Biofluidmechanik, Humboldt Universität, Deutschland. deibris@hotmail.com
Abstract:
Objective of the present study is the hemodynamic investigation of the fluidmechanics a ball valve developed for ventricular assist devices with improved parameters. The flow was investigated using Digital Particle Image Velocimetry with an enlarged model of the valve (2.8:1). The flow was recorded using a high speed video camera with 250 fps and analyzed by a cross-correlation method implemented in the software DaVis by LaVision. The results confirmed the improvement of the ball valve performed by a numerical study. The flow separations on the ball valve housing observed in the first design version was eliminated in the final design of the novel valve. Hence a valve of this design applied in ventricular assist devices can be expected to have a lower rate of thromboembolic complications.
More Related Videos
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...