Related Experiment Video
Updated: Aug 19, 2026

Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
Published on: December 2, 2014
Aortic elasticity measurement using ECG-gated multi-slice computed tomography
J M Boese1, M Ganten, D Leitermann
1Deutsches Krebsforschungszentrum Heidelberg, Deutschland. j.boese@dkfz.de
Abstract:
Vessel elasticity is a functional parameter of the vessel wall which is correlated with cardiovascular disease. Non-invasive measurement of elasticity is difficult. The aim of this project was to develop a method to measure aortic elasticity using multi-slice computed tomography (CT). The method is based on the reconstruction of time-resolved images of the aortic cross-section from CT raw data and ECG data. It can be integrated into a standard CT angiography examination without the need for additional radiation dose. The method was validated using a phantom set-up with porcine aortic specimens. Results show a good agreement between an optical reference and the CT measurement. Patient examinations show that changes in aortic wall distension can be measured and that these changes are reduced in patients with atherosclerosis.
More Related Videos
09:32Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Aortic Regurgitation II: Clinical Features and Diagnostic Tests