Detection of imported malaria with the Cell-Dyn 4000 hematology analyzer
Peter C Wever1, Yvonne M C Henskens, Piet A Kager
1Section of Parasitology, Department of Medical Microbiology, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands. p.c.wever@amc.uva.nl
Abstract:
The sensitivity and specificity of the Cell-Dyn 4000 hematology analyzer in the diagnosis of imported malaria were studied with samples from patients in an academic hospital setting. The performance of the Cell-Dyn 4000 hematology analyzer was compared with that of conventional diagnostic methods for malaria. The Cell-Dyn 4000 hematology analyzer detected hemozoin-containing depolarizing monocytes in 29 of 58 patients with malaria and 2 of 55 patients without malaria. The presence or absence of depolarizing monocytes in patients with malaria was related to duration of symptoms before presentation for malaria analysis. A second parameter, pseudoreticulocytosis due to nuclear material of intraerythrocytic malaria parasites, was detected by the Cell-Dyn 4000 hematology analyzer almost exclusively in Plasmodium falciparum malaria patients with parasitemia levels of >/=0.5%. Attention to these abnormalities in medical centers without tropical disease expertise may decrease a delay in the diagnosis of malaria.
More Related Videos
10:50Detection and Quantification of Plasmodium falciparum in Aqueous Red Blood Cells by Attenuated Total Reflection Infrared Spectroscopy and Multivariate Data Analysis
Published on: November 2, 2018
07:04Detection of Plasmodium Sporozoites in Anopheles Mosquitoes using an Enzyme-linked Immunosorbent Assay
Published on: September 30, 2021