Real-time pulse oximetry artifact annotation on computerized anaesthetic records
Richard Karl Gostt1, Graeme Dennis Rathbone, Adam Paul Tucker
1Department of Electronic Engineering, La Trobe University, Bundoora, Victoria 3083, Australia. RichardGostt@gmx.net
Objectives:
Adoption of computerised anaesthesia record keeping systems has been limited by the concern that they record artifactual data and accurate data indiscriminately. Data resulting from artifacts does not reflect the patient's true condition and presents a problem in later analysis of the record, with associated medico-legal implications. This study developed an algorithm to automatically annotate pulse oximetry artifacts and sought to evaluate the algorithm's accuracy in routine surgical procedures.
Methods:
MacAnaesthetist is a semi-automatic anaesthetic record keeping system developed for the Apple Macintosh computer, which incorporated an algorithm designed to automatically detect pulse oximetry artifacts. The algorithm labeled artifactual oxygen saturation values < 90%. This was done in real-time by analyzing physiological data captured from a Datex AS/3 Anaesthesia Monitor. An observational study was conducted to evaluate the accuracy of the algorithm during routine surgical procedures (n = 20). An anaesthetic record was made by an anaesthetist using the Datex AS/3 record keeper, while a second anaesthetic record was produced in parallel using MacAnaesthetist. A copy of the Datex AS/3 record was kept for later review by a group of anaesthetists (n = 20), who judged oxygen saturation values < 90% to be either genuine or artifact.
Results:
MacAnaesthetist correctly labeled 12 out of 13 oxygen saturations < 90% (92.3% accuracy). A post-operative review of the Datex AS/3 anaesthetic records (n = 8) by twenty anaesthetists resulted in 127 correct responses out of total of 200 (63.5% accuracy). The remaining Datex AS/3 records (n = 12) were not reviewed, as they did not contain any oxygen saturations <90%.
Conclusions:
The real-time artifact detection algorithm developed in this study was more accurate than anaesthetists who post-operatively reviewed records produced by an existing computerised anaesthesia record keeping system. Algorithms have the potential to more accurately identify and annotate artifacts on computerised anaesthetic records, assisting clinicians to more correctly interpret abnormal data.
More Related Videos
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse
The pulse serves as a clinical indicator...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...