Expulsions following 1000 GyneFix insertions
1Abacus Centres for Contraception and Reproductive Health, North Mersey Community NHS Trust, Liverpool, UK.
The Journal of Family Planning and Reproductive Health Care
|November 30, 2002
Summary
The GyneFix intrauterine device expulsion rate in clinical use is 7.6%, higher than in trials. Early expulsion may link to insertion technique, while late expulsion reasons remain unclear.
Area of Science:
- Reproductive Health
- Contraception
- Gynecology
Background:
- The GyneFix intrauterine device (IUD) has been utilized in family planning services since 1997.
- Clinical trials suggest a low expulsion rate for the GyneFix IUD.
Purpose of the Study:
- To determine the expulsion rates of the GyneFix IUD in routine clinical practice.
- To investigate potential factors contributing to IUD expulsion.
Main Methods:
- Retrospective analysis of 1000 GyneFix IUD insertions.
- Inclusion of client parity and clinician experience as variables.
- Opportunistic client consultations were conducted.
Main Results:
- The overall expulsion rate was 7.6%, exceeding rates reported in clinical trials.
- Most expulsions (4.7%) occurred within 3 months of insertion, potentially linked to insertion technique.
- Late expulsions occurred up to 28 months post-insertion; clinician experience correlated with lower late expulsion rates.
- Four unplanned pregnancies resulted from unnoticed expulsions.
Conclusions:
- The GyneFix IUD expulsion rate in this service is higher than previously reported.
- Early expulsions may be associated with suboptimal anchoring of the device.
- Further investigation into the causes of late expulsions is warranted.
- Recommendations include patient education on device checking and confirmation of unnoticed expulsions via imaging.