Abstract:
A sodium montmorillonite (SWy-2)-modified carbon-paste electrode has been examined for determination of trace levels of mercury. Because of its strong cation-exchange and adsorptive characteristics, SWy-2 greatly improves the sensitivity of determination of Hg(2+). Hg(2+) is preconcentrated and reduced on the modified electrode surface at -0.40 V and then stripped from the electrode surface during the positive potential sweep. The conditions used for determination, e.g. supporting electrolyte, pH, amount of SWy-2, accumulation potential, and accumulation time, were optimized. The peak current was linearly dependent on the concentration of mercury from 1 x 10(-9) to 5 x 10(-7) mol L(-1). The detection limit (signal-to-noise ratio=3) was 1 x 10(-10) mol L(-1) after accumulation for 6 min. When the SWy-2-modified carbon-paste electrode was used to detect mercury in water samples the average recovery was 101.11%.
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Electrodeposition
Electrodeposition can...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Controlled-Potential Coulometry: Electrolytic Methods