Flat-panel digital mammography system: contrast-detail comparison between screen-film radiographs and hard-copy
Sankararaman Suryanarayanan1, Andrew Karellas, Srinivasan Vedantham
1Departments of Radiology, University of Massachusetts Medical School, Worcester, Mass, USA.
Purpose:
To compare the contrast-detail (CD) characteristics of screen-film (SF) and postprocessed digital images by using a phantom-based method.
Materials And Methods:
Images of a CD phantom with polymerized methyl methacrylate were acquired with SF and full-field digital mammography systems at matched exposure conditions. A four-alternative forced-choice experiment was conducted with seven observers participating in the study. Each observer was required to identify randomly located disks in phantom images from which detection curves were computed. The CD diagrams for the SF and digital systems were estimated from the detection curves and compared at 50% and 62.5% threshold levels. Furthermore, a theoretic model was used to estimate the CD performance of the SF and digital systems.
Results:
Analysis of covariance for mixed models was used with the natural logarithm of disk thickness as the dependent variable, the natural logarithm of disk diameter as the covariate, and the observer as a random factor. The results of statistical analysis indicated significant differences between the CD characteristics of SF and digital mammographic images at both 50% (P <.001) and 62.5% (P <.001) detection thresholds.
Conclusion:
The authors conclude that digital CD curves, on average, exhibit threshold contrast characteristics that are lower (better) than those of SF mammography.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT