Related Experiment Videos
[Heart assist systems with integrated microsensors (HIM 2)].
1Impella Cardiotechnik AG, Aachen, Deutschland.
Biomedizinische Technik. Biomedical Engineering
|December 6, 2002
Summary
A new 4 mm intracardiac pump system is developed for coronary surgery patients, utilizing Moulded Interconnected Device (MID) technology for a smaller integrated pressure sensor. This innovation enables transfemoral placement, improving device suitability.
Area of Science:
- Biomedical Engineering
- Medical Device Technology
- Cardiovascular Support Systems
Context:
- Current intracardiac pump systems offer physiological support for coronary surgery but lack transfemoral suitability due to size.
- Existing 6.5 mm micro blood pumps with integrated sensors are too large for transfemoral device placement.
Purpose:
- To develop and test a significantly smaller intracardiac pump (4 mm outer diameter) suitable for transfemoral placement.
- To apply Moulded Interconnected Device (MID) technology for the first time in a medical device to integrate a miniaturized pressure sensor.
Summary:
- A novel 4 mm (12F) intracardiac pump is being developed, incorporating a miniaturized pressure sensor using Moulded Interconnected Device (MID) technology.
- MID technology is employed to integrate a smaller pressure sensor onto the blood pump, addressing packaging and connection challenges.
- This advancement aims to facilitate transfemoral placement of the intracardiac pump system.
Impact:
- Enables the development of smaller, transfemoral-compatible intracardiac pumps for enhanced patient support during coronary surgery.
- Paves the way for novel packaging and connection solutions for miniaturized sensors in medical devices.
- Advances the application of MID technology in the medical field for next-generation cardiovascular support.