Related Experiment Videos
Ultrasound properties of human prostate tissue during heating
A E Worthington1, J Trachtenberg, M D Sherar
1Ontario Cancer Institute/Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada.
Abstract:
Changes in the ultrasound (US) properties of tissue during heating affect the delivery of US thermal therapy and may provide a basis for US image monitoring of thermal therapy. The US attenuation coefficient and backscatter power of fresh human prostate tissue were measured as the tissue was heated. Samples of human prostate were obtained directly from autopsies and heated rapidly to final temperatures of 45 degrees C, 50 degrees C, 55 degrees C, 60 degrees C and 65 degrees C. A 5.0-MHz transducer was scanned in a raster pattern over the tissue and radiofrequency (RF) data were collected at 36 uncorrelated positions. Both attenuation and backscatter were measured over the frequency range 3.5 to 7.0 MHz at each min of a 30-min heating. Little change was observed in attenuation or backscatter at 55 degrees C or less. The attenuation coefficient and backscatter power increased by factors of 1.25 and 5, respectively, during the 60 degrees C heating. During the 65 degrees C heating, the same properties showed increases by factors of 2.7 and 9.