Comparison of a sidestream capnograph and a mainstream capnograph in mechanically ventilated dogs
Francisco J Teixeira Neto1, Adriano B Carregaro, Rodrigo Mannarino
1Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista, Botucatu, SP, 18618-000, Brazil.
Objective:
To compare the ability of a sidestream capnograph and a mainstream capnograph to measure end-tidal CO2 (ETCO2) and provide accurate estimates of PaCO2 in mechanically ventilated dogs.
Design:
Randomized, double Latin square.
Animals:
6 healthy adult dogs.
Procedure:
Anesthesia was induced and neuromuscular blockade achieved by IV administration of pancuronium bromide. Mechanical ventilation was used to induce conditions of standard ventilation, hyperventilation, and hypoventilation. While tidal volume was held constant, changes in minute volume ventilation and PaCO2 were made by changing the respiratory rate. Arterial blood gas analysis was performed and ETCO2 measurements were obtained by use of either a mainstream or a sidestream capnographic analyzer.
Results:
A linear regression model and bias analysis were used to compare PaCO2 and ETCO2 measurements; ETCO2 measurements obtained by both capnographs correlated well with PaCO2. Compared with PaCO2, mainstream ETCO2 values differed by 3.15 +/- 4.89 mm Hg (mean bias +/- SD), whereas the bias observed with the sidestream ETCO2 system was significantly higher (5.65 +/- 5.57 mm Hg). Regardless of the device used to measure ETCO2, bias increased as PaCO2 exceeded 60 mm Hg.
Conclusions And Clinical Relevance:
RelevancehAlthough the mainstream cas slightly more accurate, both methods of ETCO2 measurement correlated well with PaCO2 and reflected changes in the ventilatory status. However, ETCO2 values > 45 mm Hg may inaccurately reflect the severity of hypoventilation as PaCO2 may be underestimated during conditions of hypercapnia (PaCO2 > 60 mm Hg).
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management