Evaluation of a discrete-sample computer-directed clinical analyzer.

R L Forrester, W Collinge, P Hashimoto

    Clinical Chemistry
    |February 1, 1976
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Early post-operative diet upgrade in older patients may improve energy and protein intake but patients still eat poorly: an observational pilot study.

    Journal of human nutrition and dietetics : the official journal of the British Dietetic Association·2018
    Same author

    Cost-Effectiveness Analysis of Testing for Brca Mutations in Women Diagnosed with Ovarian Cancer and their Female First-Degree Relatives: A Uk Health Service Perspective.

    Value in health : the journal of the International Society for Pharmacoeconomics and Outcomes Research·2016
    Same author

    Kinship care of the abused child: the New Zealand experience.

    Child welfare·2001
    Same author

    Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles.

    Journal of biomedical materials research·1998
    Same author

    Effect of diabetes and peritonitis on the peritoneal equilibration test.

    Kidney international·1995
    Same author

    An effect of gravity on bud-burst in balsam fir.

    Tree physiology·1986

    The Gilford System 3500 demonstrates excellent accuracy and precision for multiple clinical chemistry tests, offering a reliable and easy-to-operate solution for laboratories. Its performance meets clinical acceptance criteria.

    Area of Science:

    • Clinical Chemistry
    • Laboratory Instrumentation
    • Diagnostic Assays

    Background:

    • The Gilford System 3500 is a widely used automated clinical chemistry analyzer.
    • Evaluating the performance of laboratory instruments is crucial for ensuring reliable patient results.

    Purpose of the Study:

    • To assess the operational characteristics of the Gilford System 3500 over a six-month period.
    • To determine the accuracy, precision, carry-over, reliability, and ease of operation of the instrument.

    Main Methods:

    • Accuracy was determined by comparing results to established manual methods, calculating correlation coefficients for eight key analytes.
    • Within-run and day-to-day precision were evaluated using commercial calibration sera at varying concentrations.
    • Carry-over, reliability, and ease of operation were also assessed.

    Related Experiment Videos

    Main Results:

    • High correlation coefficients were observed for all tested analytes, with values ranging from 0.96 to 0.99.
    • Within-run precision (CV) values were consistently low across different analyte concentrations.
    • Day-to-day precision met accepted clinical criteria, and carry-over was less than 1% for all tests.
    • Instrument reliability was excellent, and training time was minimal.

    Conclusions:

    • The Gilford System 3500 is a highly accurate and precise instrument for a broad range of clinical chemistry tests.
    • The system is reliable, easy to operate, and demonstrates flexibility, making it a valuable tool for clinical laboratories.