Related Experiment Video
Updated: Sep 28, 2026

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Hidden errors of aneroid sphygmomanometers
Jason J S Waugh1, Manesh Gupta, Julie Rushbrook
1Department of Obstetrics and Gynaecology, Leicester-Warwick Medical School, Leicester Royal Infirmary, Leicester, UK.
Background:
Measurement of blood pressure remains the most commonly performed screening test in medical practice. With the likely removal of mercury sphygmomanometers from the workplace alternative devices are required. Of these the aneroid sphygmomanometer is popular both in the community and hospital setting. We investigated the accuracy of all the aneroid and mercury sphygmomanometers during dynamic calibration within a tertiary referral maternity hospital.
Methods:
We compared the accuracy of 39 aneroid and 36 mercury sphygmomanometers to a recently calibrated and serviced mercury sphygmomanometer (the accepted gold standard). All devices were in current clinical use. Using three blinded, trained observers, 30 different pressures were checked throughout the pressure range following British Hypertension Society protocol guidelines.
Results:
Only 31 (86%) of the mercury devices and 36 (92%) of the aneroid devices were in adequate working condition and suitable for analysis. Significantly more aneroid devices had systematic errors of > 5 mmHg (19 versus 3%, < 0.05). Fifty percent of aneroid devices had at least one reading > 10 mmHg out compared to only 10% of mercury devices (chi square programme).
Conclusions:
Aneroid sphygmomanometers in apparent good working order are inaccurate compared to mercury devices. Some of these faults can only be detected during dynamic testing. To minimize the risk of erroneous blood pressure recording, aneroid devices should be regularly checked for accuracy using dynamic calibration methods as recommended in validation protocols.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure

