Related Experiment Video
Updated: Sep 28, 2026

Programmed Electrical Stimulation in Mice
Published on: May 26, 2010
Implantable cardioverter defibrillator proarrhythmia due to an interaction with "noncompetitive atrial pacing": an
Thomas A Cron1, Beat Schaer, Stefan Osswald
1Division of Cardiology, University Hospital, Petersgraben 4, CH-4031 Basel, Switzerland. tcron@uhbs.ch
Abstract:
An episode of device related proarrhythmia observed with a dual chamber ICD is reported. Ventricular tachycardia was induced due to an inappropriate interaction of programmed device settings with automatic algorithms, including noncompetitive atrial pacing (NCAP), an algorithm used to prevent atrial tachyarrhythmias.