Related Experiment Video
Updated: Feb 18, 2026

Low-cost Custom Fabrication and Mode-locked Operation of an All-normal-dispersion Femtosecond Fiber Laser for Multiphoton Microscopy
Published on: November 22, 2019
A beam-splitting device for use with fiber-coupled laser light sources for photodynamic therapy
Leroy M Wood1, David A Bellnier, Allan R Oseroff
1The Photodynamic Therapy Center, Roswell Park Cancer Institute, Buffalo, NY 14263, USA.
Abstract:
A device that divides light into eight, four or two beams of equivalent power with only minor total power loss was designed, built and tested. The apparatus accepts light from a 200 microm diameter, 0.16 numerical aperture, silica-silica multimode optical fiber connected to one of several laser light sources for photodynamic therapy (PDT) of cancer. The incorporation of a variable iris diaphragm into the optical couplers allows the power of the beams to be independently set. Each of the beams can be coupled to a 400 or 600 microm diameter optical fiber to deliver the therapeutic light to the patient. This device is used in our institute for PDT of patients with either numerous small malignant tumors or single tumors with large surface area.