[Lateral radiography of the knee with single-leg standing]
Kan Ohmori1, Tsuyoshi Sugawara, Katsuyoshi Murakami
1Center for Radiological Sciences, Iwate Medical University Hospital, Japan.
Abstract:
The purpose of this investigation was to accomplish reproducible radiography of single-leg standing lateral radiography of the knee by adjusting lateral rotation using a ruler to measure foot position. After preliminary assessment of three-dimensional CT of the knees of normal volunteers, the best adjustment of external rotation was estimated. A ruler was made for use in adjusting the angle of knee rotation by measuring foot rotation. Based on the foot rotation measured by this ruler, the positioning of radiography was adjusted to correct rotation. Rotation was estimated by the distance between the posterior edges of the lateral and medial femoral condyles. Fifteen-degree and 17.5-degree rotations were used for correction. Correction of rotation was 17 degrees on average. This helped not only to correct external rotation in the initial radiography but also to correct rotation for repeat radiography. Our method is quantitative and highly reproducible, and it increases the success rate of lateral knee radiography.