Related Experiment Video
Updated: Sep 27, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
End-expiratory lung impedance change enables bedside monitoring of end-expiratory lung volume change
1Department of Anaesthesiology, Emergency and Intensive Care Medicine, University of Göttingen, Robert-Koch-Str 40, 37075 Göttingen, Germany. mail@josehinz.de
Objective:
The aim of the study was to investigate the effect of lung volume changes on end-expiratory lung impedance change (ELIC) in mechanically ventilated patients, since we hypothesized that ELIC may be a suitable parameter to monitor lung volume change at the bedside.
Design:
Clinical trial on patients requiring mechanical ventilation.
Settings:
Intensive care units of a university hospital.
Patients:
Ten mechanically ventilated patients were included in the study.
Intervention:
Patients were ventilated in volume-controlled mode with constant flow and respiratory rate. In order to induce changes in the end-expiratory lung volume (EELV), PEEP levels were increased from 0 mbar to 5 mbar, 10 mbar, and 15 mbar. At each PEEP level EELV was measured by an open-circuit nitrogen washout manoeuvre and ELIC was measured simultaneously using Electrical Impedance Tomography (EIT) with sixteen electrodes placed on the circumference of the thorax and connected with an EIT device. Cross-sectional electro-tomographic measurements of the thorax were performed at each PEEP level, and a modified Sheffield back-projection was used to reconstruct images of the lung impedance. ELIC was calculated as the average of the end-expiratory lung impedance change. RESULTS. Increasing PEEP stepwise from 0 mbar to 15 mbar resulted in an linear increase of EELV and ELIC according to the equation: y =0.98 x -0.68, r(2)=0.95.
Conclusion:
EIT is a simple bedside technique which enables monitor lung volume changes during ventilatory manoeuvres such as PEEP changes.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...