The electrical dynamics of laparoscopic sterilization.

T Engel, F W Harris

    The Journal of Reproductive Medicine
    |July 1, 1975
    PubMed
    Summary

    Preventing bowel burns during laparoscopic sterilization requires low voltage. Coagulating fallopian tubes needs only 100 volts peak, while charring requires 700 volts peak.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Impact of NNLO QED corrections on lepton-proton scattering at MUSE.

    The European physical journal. A, Hadrons and nuclei·2023
    Same author

    Refining the dermatological spectrum in primary immunodeficiency: mucosa-associated lymphoid tissue lymphoma translocation protein 1 deficiency mimicking Netherton/Omenn syndromes.

    The British journal of dermatology·2019
    Same author

    Re-phrasing the question: A simple tool for evaluation of adherence to therapy in patients with inflammatory bowel disease.

    United European gastroenterology journal·2017
    Same author

    A multi-centre prospective evaluation of the Check-Direct ESBL Screen for BD MAX as a rapid molecular screening method for extended-spectrum beta-lactamase-producing Enterobacteriaceae rectal carriage.

    The Journal of hospital infection·2017
    Same author

    Addition of an immunomodulator can reverse antibody formation and loss of response in patients treated with adalimumab.

    Alimentary pharmacology & therapeutics·2016
    Same author

    Driving while using a smartphone-based mobility application: Evaluating the impact of three multi-choice user interfaces on visual-manual distraction.

    Applied ergonomics·2016

    Area of Science:

    • Minimally Invasive Surgery
    • Electrosurgery Principles

    Background:

    • Laparoscopic sterilization is a common procedure.
    • Bowel burns are a potential complication of electrosurgery.

    Purpose of the Study:

    • Determine the minimum voltage for safe laparoscopic sterilization.
    • Investigate electrosurgical parameters to prevent bowel burns.

    Main Methods:

    • Measured electrical resistance of tissues during coagulation.
    • Utilized a "CUT" waveform at varying voltage levels.
    • Evaluated electrosurgical effects on sample tissues and during sterilization.

    Main Results:

    • Coagulation of fallopian tubes is achievable at 100 volts peak.
    • Electrosurgical cutting or fulguration requires approximately 700 volts peak.
    • Tissue electrical resistance changes were documented.

    Conclusions:

    • Low voltage electrosurgery is effective for laparoscopic sterilization.
    • Understanding voltage requirements can minimize risks like bowel burns.
    • Basic electrosurgical principles are crucial for surgical safety.