In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and
Bradley L Fricke1, Lane F Donnelly, Donald P Frush
1Departments of Radiology and Pediatrics, Cincinnati Children's Hospital Medical Center and University of Cincinnati College of Medicine, 3333 Burnet Ave., Cincinnati, OH 45229-3039, USA.
Objective:
The purpose of our study was to evaluate the amount of radiation dose reduction and its effect on image quality when using an in-plane bismuth breast shield for multidetector CT (MDCT) of the chest and abdomen in female pediatric patients.
Subjects And Methods:
Fifty consecutive MDCT examinations (chest, 29; abdomen, 21) of female pediatric patients (mean age, 9 years; range, 2 months-18 years) were performed with a 2-ply (1.7 g of bismuth per square centimeter) bismuth shield (three sizes to accommodate patients of varying sizes) overlying the patient's breasts. MDCT images were evaluated for a perceptible difference in image quality in the lungs at the anatomic level under the shield as compared with nonshielded lung and whether the images were of diagnostic quality. In addition, 2-mm regions of interest were placed in the peripheral anterior and posterior portions of each lung in shielded and nonshielded areas, and noise (standard deviation in Hounsfield units) was measured in the regions. Differences among the regions in noise were compared for shielded versus nonshielded areas (paired t test). To measure differences in actual dose, we also evaluated the breast shield with an infant anthropomorphic phantom using thermoluminescent detectors in the breast tissue. The phantom was imaged with and without the breast shield using identical MDCT parameters.
Results:
All MDCT scans of patients were of diagnostic quality with no perceptible difference in image quality in shielded versus nonshielded lung. We found no statistically significant difference in noise between the shielded and nonshielded lung regions of interest (shielded: mean noise, 17.3 H; nonshielded: mean noise, 18.8 H; p = 0.5180). Phantom measurements revealed a 29% reduction in radiation dose to the breast when a medium-dose MDCT protocol was used.
Conclusion:
Bismuth in-plane breast shielding for pediatric MDCT decreased radiation dose to the breast without qualitative or quantitative changes in image quality.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...