Comparison of computed radiography and film/screen combination using a contrast-detail phantom
Z F Lu1, E L Nickoloff, J C So
1Department of Radiology, Columbia University, MHB 3-265, 177 Ft. Washington Avenue, New York, New York 10032, USA. zfl1@columbia.edu
Abstract:
The purpose of this work is to compare computed radiography (Kodak CR 400) and film/screen combination (Speed 400) systems in regards of patient dose, technique settings, and contrast-detail detectability. A special contrast-detail phantom with drilled holes of varying diameter (detail) and varying depth (contrast) was utilized. Various thicknesses of the Lucite sheets were utilized to simulate scattering tissues. Images of the phantom were acquired using a range of 60-120 kVp for film/screen and CR with a conventional x-ray tube and then for CR with additional 2 mm aluminum added filtration to the x-ray beam. The patient entrance skin dose was measured while maintaining 1.6 o.d. for film/screen images and 1900 Exposure Index for CR images. CR phantom images were displayed on the diagnostic workstation for soft copy reading as well as printed on films for hard copy reading on viewbox. Four physicists evaluated the images by scoring the threshold target depth along the row of the same target diameter. Detection ratio was calculated by counting the number of detectable targets divided by the total number of targets in the phantom. The overall score was related to the patient entrance skin dose, kVp, and the thickness of the scattering material. The patient entrance skin dose was reduced as the additional aluminum filter was added to the x-ray beam. Our findings suggested using a higher kVp setting and additional added filtration would reduce the patient entrance skin dose without compromising the contrast-detail detectability, which was compensated by the contrast manipulation on soft-copy display workstations.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography