Related Experiment Video
Updated: Sep 21, 2026

Electrophoretic Separation of Proteins
Published on: June 12, 2008
[Study on the separation of proteins in human serum by high performance capillary zone electrophoresis]
Abstract:
A high performance capillary zone electrophoretic method has been developed for the separation of proteins in human serum. Sera were diluted with a buffer (0.05 mol/L boric acid, pH 8.80), and then introduced hydrodynamically into a 37 cm (32 cm to detector) x 50 microns i.d. fused-silica capillary, electrophoresed for 12 min using a boric buffer solution (0.10 mol/L, pH 9.35; containing 4 g/L of polyethylene glycol 8000), and detected at the wavelength of 200 nm. The method is simple, rapid and repeatable.
Related Concept Videos
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such as cells...
Ion-Exchange Chromatography
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...