Related Experiment Videos
[Analysis of a novel platinum anticancer compound HJ5 by high performance liquid chromatography]
Yang Liu1, Jian He, Zhu-dong Liu
1Research Center of Precious Metal-Based Drugs, Kunming Institute of Precious Metals, Kunming 650221, China.
Abstract:
Cis-dichloroamine (4-amino-2, 2, 6, 6-tetramethylpiperidine-1-oxyl) platinum (II) (abbreviated to HJ5) is a novel platinum anticancer compound which has prospects for research and application. In order to establish a conventional analytical method, an HPLC has been studied. Under the conditions of acetonitrile-water (10:90, volume ratio) as mobile phase, detection wavelength of 210 nm, a Phenomenex column C18(150 mm x 3.9 mm i.d., 10 microns). The linear equation, recovery and accuracy of the method were determined with external standard method. The HPLC has satisfactory resolution between the peaks of HJ5 and impurities. The peak areas of HJ5 were linear to its amounts detected and the detection limit was 0.07 microgram. The hydration of HJ5 can be inhibited effectively by NaCl solution. In a 9 g/L NaCl solution, the stability can be kept for more than 2 hours which is enough for analysis.