Juha-Pekka Nuutinen1, Tero Välimaa, Claude Clerc
1Institute of Biomaterials, Tampere University of Technology, PO Box 589, FIN-33101, Tampere, Finland.
Abstract:
The aim of this study was to characterize the mechanical properties and in vitro degradation of bioresorbable knitted stents. Each stent was knitted using a single self-reinforced fibre made out of either PLLA or 96L/4D PLA or 80L/20G PLGA. The mechanical and physical properties of the fibres and stents were measured before and after gamma sterilization, as well as during in vitro degradation. The mechanical properties of the knitted stents made out of bioresorbable fibres were similar to those of commercially available metallic stents. The knitting geometry (loop height) had a marked effect on the mechanical properties of the stents. The rate of in vitro degradation in mechanical and physical properties for the PLLA and 96L/4D PLA stents was similar and significantly lower than that of the 80L/20G PLGA stents. The 80L/20G PLGA stents lost about 35% of their initial weight at 11 weeks. At this time, they had lost all their compression resistance strength. These data can be used as a guideline in planning further studies in vivo.