Spontaneously breathing lung model comparison of work of breathing between automatic tube compensation and pressure
Yuji Fujino1, Akinori Uchiyama, Takashi Mashimo
1Intensive Care Unit, Osaka University Hospital, Suita, Japan. fujino@hp-icu.med.osaka-u.ac.jp.
Introduction:
Automatic tube compensation (ATC), a new ventilation mode that compensates for the work of breathing imposed by endotracheal tube resistance, recently became commercially available.
Methods:
We conducted a laboratory study with a lung model and a Nellcor Puritan Bennett 840 ventilator to compare ATC and pressure-support ventilation (PSV). A bellows-in-a-box lung model simulated spontaneous breathing with the following settings: respiratory rate 10 breaths/min, inspiratory time 1.0 s, peak inspiratory flow 60 L/min without connecting to the ventilator and endotracheal tube (ETT). At each ETT size (5, 6, 7, 8 and 9 mm inner diameter) 100% ATC was compared with pressure support (PS) of 0, 2, 4, 6, 8, and 10 cm H(2)O at positive end-expiratory pressure (PEEP) of 0 and 5 cm H(2)O. The negative deflection (PI) of the "alveolar" pressure (ie, pressure inside the bellows, P(alv)) and the delay time were measured. The PI and total pressure-time product (PTP(tot)) integrated from P(alv) were analyzed. PTP(tot) was subdivided into PTP(trig) (the PTP from the beginning of inspiration to the minimum P(alv)) and PTP(supp) (the PTP from the minimum P(alv) to the return to baseline P(alv)).
Results:
At PEEP of 0 cm H(2)O: with ETTs of 5, 6, and 7 mm the PI values with ATC corresponded to PS of 0-4 cm H(2)O; with the 8-mm ETT the PI values corresponded to PS of 0 cm H(2)O; with the 9-mm ETT the PI values corresponded to PS of 0-2 cm H(2)O. At PEEP of 5 cm H(2)O, with all ETT sizes the PI values corresponded to PS of 0 cm H(2)O. PTP(tot) and PTP(supp) of ATC corresponded to: PS of 2-4 cm H(2)O with the 5-mm ETT; PS of 2 cm H(2)O with the 6-mm ETT; PS of 0-2 cm H(2)O with the 7-mm ETT; and PS of 0 cm H(2)O with the 8- and 9-mm ETTs, at PEEP of 0 cm H(2)O. PEEP of 5 cm H(2)O was not tested for PTP. PTP(trig) with ATC showed comparable or greater values with each size of ETT.
Conclusions:
ATC with a Nellcor Puritan Bennett 840 ventilator provided inspiratory ventilatory support corresponding to PS of
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Application of Integration: Problem Solving
Mechanism of Breathing I: Inspiration
The respiratory system, an essential network for breathing, comprises the conducting and respiratory zones, each playing a crucial role in the overall process of respiration. Let us explore the detailed mechanism of inspiration, or inhalation, which is the first phase of the respiratory cycle.
Pathway of Air during Inspiration
During inspiration, air enters our body through the nose or mouth and moves through the conducting zone,...