Related Experiment Videos
Determination of benzene in urine by static headspace gas chromatography
Ljiljana Skender1, Irena Brcić, Silva Zuzul
1Institute for Medical Research and Occupational Health, Zagreb, Croatia. Ljiljana.Skender@imi.hr
Abstract:
This paper describes the application of static headspace gas chromatography in determining benzene in urine. The method was analytically validated for sensitivity (DL = 42 ng/l), repeatability (RSD = 3% and 4%), and accuracy (71%), and was applied in measuring urine benzene in nonsmokers (N = 14) and smokers (N = 18). All urine samples had measurable benzene concentrations. The method proved sensitive enough to establish a significant statistical difference (P < 0.000614) in urine benzene concentrations between smokers (mean = 760; range = 181-1869 ng/l) and nonsmokers (mean = 214; range = 61-515 ng/l).