Both the OxyArm and Capnoxygen mask provide clinically useful capnographic monitoring capability in volunteers
James Paul1, Elizabeth Ling, Julius Hajgato
1Department of Anesthesia McMaster University Southmedic Inc Barrie, Ontario, Canada. paulj@quickclic.net
Purpose:
To compare the capnography monitoring performance of the new OxyArm (OA) with the Capnoxygen mask (CM), a conventional oxygen mask with a carbon dioxide sampling port.
Methods:
Eleven healthy volunteer adult subjects underwent capnographic monitoring (in a non-randomized, un-blinded crossover study) at baseline and while receiving oxygen at seven different flow rates (0.5, 1.0, 2.0, 4.0, 6.0, 8.0, and 10 L x min(-1)), applied first with the CM and then with the OA.
Results:
Both the OA and CM produced acceptable capnographs with consistent waveforms. The measured end-tidal (ET) CO(2) was equivalent for the two devices at all seven oxygen flow rates. On average, the ETCO(2) measured with the OA was about 2 mmHg greater than that of the CM. Regression analysis showed an inverse relationship between oxygen therapy flow rate and measured ETCO(2) whereby the measured value of CO(2) decreased as the oxygen flow rate was increased (P < 0.001). Both the CM and OA produced consistent measurements of ETCO(2) as illustrated by their reliability coefficients, 0.95 and 0.86 respectively. The biggest source of variation in measured CO(2) for both devices was inter-subject differences, followed by variable oxygen flow rates.
Conclusions:
This study suggests that the OA and CM can prove useful for respiratory monitoring and oxygen delivery in spontaneously breathing volunteers, and the OA could potentially be used as an alternative to the conventional methods of oxygen delivery and CO(2) sampling in patients.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.