Amount of air infused to patient increases as fluid flow rates decrease when using the Hotline HL-90 fluid warmer.

S Woon1, P Talke

  • 1University of California, Los Angeles 94143-0648, USA.

Journal of Clinical Monitoring and Computing
|February 6, 2003
PubMed
Summary

Fluid warming devices can infuse air into patients, increasing the risk of air embolism. Lower infusion rates significantly increase the volume of air delivered, posing a potential patient safety hazard during surgery.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Body image and psychological distress in nipple-sparing mastectomy: the roles of self-compassion and appearance investment.

Psycho-oncology·2016
Same author

Effect of pain on autonomic nervous system indices derived from photoplethysmography in healthy volunteers.

British journal of anaesthesia·2012
Same author

Photoplethysmographic waveform: what do the numbers mean?

Acta anaesthesiologica Scandinavica·2010
Same author

Effects of nitric oxide synthase inhibition on dexmedetomidine-induced vasoconstriction in healthy human volunteers.

British journal of anaesthesia·2008
Same author

Effects of intrathecally administered dexmedetomidine, MPV-2426 and tizanidine on EMG in rats.

Acta anaesthesiologica Scandinavica·2003
Same author

The hemodynamic and adrenergic effects of perioperative dexmedetomidine infusion after vascular surgery.

Anesthesia and analgesia·2000

Area of Science:

  • Anesthesiology
  • Medical Devices
  • Patient Safety

Background:

  • Fluid warming devices are recommended to prevent perioperative hypothermia.
  • Potential risks associated with these devices, such as air infusion, require investigation.

Purpose of the Study:

  • To quantify the volume of air infused from a warmed intravenous solution using a specific fluid warmer.
  • To assess the risk of air embolism associated with intraoperative fluid warming.

Main Methods:

  • Tested the HL-90 Hotline fluid warmer with L-70 tubing at various flow rates (150-3400 ml/h).
  • Collected and measured air volume from infused lactated Ringer's solution using a bubble trap.
  • Evaluated the efficacy of the L-10 Gas Vent in reducing air infusion.

Related Experiment Videos

Main Results:

  • Air volume infused significantly increased as flow rate decreased (p < 0.0001).
  • At 150 ml/h, air volume was 2.9 +/- 0.4 ml, compared to 1.0 +/- 0.2 ml at 3400 ml/h.
  • The L-10 gas eliminator did not effectively reduce infused air volume.

Conclusions:

  • Hotline fluid warmer use can lead to patient infusion of air.
  • This presents a potential risk of air embolism during surgical procedures.
  • Careful consideration of fluid warmer use and monitoring is warranted.