Peripapillary nerve fiber layer thickness measurement reproducibility using optical coherence tomography
Max A Villain1, David S Greenfield
1Department of Ophthalmology, University of Miami School of Medicine, Bascom Palmer Eye Institute, Miami, Florida 33418, USA.
Summary
Optical coherence tomography (OCT) shows good reproducibility for measuring retinal nerve fiber layer thickness in normal eyes. Even inexperienced operators can obtain reliable data for clinical use.
Area of Science:
- Ophthalmology
- Medical Imaging
- Neuroscience
Background:
- Retinal nerve fiber layer (RNFL) thickness is a key indicator of optic nerve health.
- Accurate RNFL measurements are crucial for diagnosing and monitoring glaucoma and other optic neuropathies.
- Optical coherence tomography (OCT) is a widely used imaging technique for assessing RNFL thickness.
Purpose of the Study:
- To evaluate the reproducibility of peripapillary retinal nerve fiber layer thickness measurements.
- To assess the reliability of quadrantic and clock hour sectors of RNFL thickness.
- To determine if inexperienced operators can achieve acceptable measurement variability using OCT.
Main Methods:
- Healthy volunteers' normal eyes were imaged using OCT by two inexperienced operators.
- Six 360-degree circular scans (3.4 mm diameter) were acquired per session.
- Measurements were taken on three separate days within a four-week period, with variance and coefficient of variation (CV) calculated.
Main Results:
- High intraoperator reproducibility was observed for both operators (mean CV for total RNFL: 5.3% and 4.33%).
- Interoperator reproducibility was good, with no statistically significant differences in variance for most RNFL parameters (P = .42 to .99).
- The nasal RNFL sector showed the highest variability (mean CV: 10.42% and 7.83%).
Conclusions:
- OCT-based peripapillary RNFL thickness assessments demonstrate good intraoperator and interoperator reproducibility.
- Inexperienced operators can generate reliable RNFL thickness data with acceptable variance.
- OCT is a reproducible method for RNFL thickness assessment in clinical practice.