Blood pressure measurement in an ambulatory setting: concordance between physician and patient self-measurement
E Vinyoles1, X Blancafort, C López-Quiñones
1La Mina Primary Care Centre, Family and Community Medicine Teaching Unit, Barcelona, Catalonia, Spain. 23561evb@comb.es
Insights
Physician and patient blood pressure (BP) measurements showed significant agreement in medical office settings. However, patient self-measurement tended to elevate systolic BP and reduce BP control rates.
Area of Science:
- Cardiology
- Hypertension Management
- Diagnostic Accuracy
Background:
- Accurate blood pressure (BP) monitoring is crucial for effective hypertension management.
- Discrepancies between physician and patient BP readings can impact treatment decisions.
- Ambulatory and self-measurement techniques are increasingly used to capture real-world BP data.
Purpose of the Study:
- To assess the concordance between physician and patient blood pressure measurements taken in a clinical setting.
- To compare BP readings obtained using a mercury sphygmomanometer versus an automated oscillometric device.
- To evaluate the impact of patient self-measurement on BP control.
Main Methods:
- A cross-sectional study involving 106 hypertensive patients.
- BP measurements were taken nine times per patient: three by a physician (mercury sphygmomanometer), three by the physician (automated oscillometer), and three by the patient (automated oscillometer).
- Intraclass correlation coefficients (ICCs) were calculated to determine measurement concordance.
Main Results:
- All three measurement methods demonstrated significant concordance (ICCs ranging from 0.64 to 0.83 for systolic/diastolic pressures, P < 0.001).
- Mean systolic BP was 136/80 mmHg (physician-mercury), 137/80 mmHg (physician-automated), and 139/80 mmHg (patient-automated).
- Blood pressure control rates were 48.1% (physician-mercury), 48.1% (physician-automated), and 36.8% (patient-automated), with patient self-measurement showing a statistically significant lower control rate (P < 0.05).
Conclusions:
- Physician and patient BP measurements in the medical office setting exhibit significant concordance.
- Patient self-measurement in the office setting may lead to an overestimation of systolic BP.
- The findings suggest that while patient self-measurement is concordant, it may negatively affect the assessment of BP control.
Abstract:
The aim of this study was to determine concordance between physician and patient blood pressure (BP) measurements in an ambulatory setting. A diagnostic intervention cross-sectional study using a convenience sample was employed. A total of 106 hypertensive patients were included in the study. Patients who were unable to perform their self-measurement or those with cardiac arrhythmia were excluded. BP was determined nine times in each subject in the medical office in a randomised order: BP was taken three times by the physician using a mercury sphygmomanometer (SPH-Hg), three times by the physician using a validated, automated oscillometer (Omron HEM 705 CP), and three times by the patient himself with the same device. The intraclass correlation coefficient was calculated. In all, 59 women and 47 men aged 65.7 (10) years were analysed. Mean BP measurements for the physician using the mercury sphygmomanometer, the physician using the Omron, and the patient using the same device were: 136 (15.8)/80 (11), 137 (17.9)/80 (10), and 139* (17.6)/80 (10) mmHg, respectively. BP control was 48.1, 48.1, and 36.8*% (*P < 0.05), respectively. Intraclass correlation coefficients for systolic/diastolic pressures were: 0.77/0.65 (physician-sphygmomanometer Hg, physician-Omron; P < 0.001), 0.75/0.64 (physician-sphygmomanometer Hg, patient-Omron, P < 0.001), and 0.83/0.83 (physician-Omron, patient-Omron; P < 0.001). In conclusion, the three types of measurement in the medical office were significantly concordant. Patient office self-measurement showed a tendency to increase systolic BP and worsen BP control.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...