Electrical impedance measurement as an endpoint detection method for routine coagulation tests.
1Haematology Department, Raigmore Hospital, Old Perth Road, Inverness IV2 3UJ, Scotland, UK. neil.spence@raigmore.scot.nhs.uk
British Journal of Biomedical Science
|February 8, 2003
Summary
This study adapted a platelet aggregometer to measure electrical impedance for prothrombin time (PT) testing. The impedance method shows promise, correlating with fibrinogen levels and offering a reproducible alternative to absorbance methods.
Area of Science:
- Clinical Chemistry
- Hematology
- Biomedical Engineering
Background:
- Prothrombin time (PT) is a critical measure of blood coagulation.
- Traditional PT measurement relies on absorbance-based methods.
- Adapting existing laboratory equipment for novel diagnostic applications is valuable.
Purpose of the Study:
- To adapt a whole-blood platelet aggregometer for measuring electrical impedance changes during PT estimations.
- To compare the impedance-based PT method with the conventional absorbance method.
- To assess the correlation between impedance signal amplitude and plasma fibrinogen concentration.
Main Methods:
- Utilized a whole-blood platelet aggregometer to monitor electrical impedance changes in plasma.
- Analyzed the impedance curve for an acceleration phase indicative of coagulation onset.
- Correlated the amplitude of impedance change with plasma fibrinogen levels.
- Performed statistical analysis comparing PT values derived from impedance and absorbance methods.
Main Results:
- The impedance curve exhibited an acceleration phase analogous to absorbance-based coagulation.
- A significant correlation was observed between impedance signal amplitude and plasma fibrinogen concentration.
- While statistically different, PT values from impedance and absorbance methods showed good correlation.
- The impedance method demonstrated reproducibility but was noted as laborious.
Conclusions:
- Electrical impedance measurement using an adapted platelet aggregometer is a viable method for PT estimation.
- The method shows potential for assessing fibrinogen concentration.
- Further optimization with more sensitive instrumentation and improved electrode design is recommended.
- Reagent system modifications may enhance the impedance-based coagulation assay.