Validation of the Natus CO-Stat End Tidal Breath Analyzer in children and adults
H J Vreman1, R J Wong, P Harmatz
1Departments of Pediatric, Stanford University School of Medicine, Stanford, CA 94305-5208, USA. henk.vreman@stanford.edu
Objective:
The performance of a point-of-care, noninvasive end tidal breath carbon monoxide analyzer (CO-Stat End Tidal Breath Analyzer, Natus Medical Inc.) that also reports end tidal carbon dioxide (ETCO2) and respiratory rate (RR), was compared to established, marketed (predicate) devices in children (n = 39) and adults (n = 48) who are normal or at-risk of elevated CO excretion.
Methods:
Concentrations of end tidal breath CO (ETCO), room air CO, ETCO corrected for inhaled CO (ETCOc), ETCO2, and RR were measured with the CO-Stat analyzer and the data compared to those obtained from the same subjects using the Vitalograph BreathCO monitor (Vitalograph, Inc.) for ETCOc and the Pryon CO2 monitor (SC210 and SC300, Pryon Corp) for ETCO2 and RR. Adults and children were studied at three medical centers. The data were analyzed by paired t-tests and linear regression. Bias and imprecision between the CO-Stat analyzer and the predicate devices was calculated by the method of Bland and Altman.
Results:
Paired t-tests, performed on the three parameters measured with the CO-Stat analyzer and predicate devices showed that only the ETCOc values in the adults and the ETCO2 values in the children were significantly different (lower, p < or = 0.0001, and higher, p < or = 0.0001, respectively). The mean bias and imprecision of the CO-Stat analyzer for adult ETCOc and children ETCO2 measurements were -0.9 +/- 1.2 ppm and 0.4 +/- 0.6%, respectively. Linear regression analysis for the ETCOc results in children and adults had a high degree of correlation (r = 0.91 and 0.98, respectively).
Conclusions:
We conclude that in a clinical environment the Natus CO-Stat End Tidal Breath Analyzer performs at least as well as predicate devices for the measurements of ETCOc, ETCO2, and RR.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...