Use of internal coils for independent and direct MR imaging-guided endovascular device tracking
Reed A Omary1, Jordin D Green, Wayne S Fang
1Department of Radiology, Northwestern University Medical School, 676 North St. Clair, Suite 800, Chicago, Illinois 60611, USA. romary@radiology.northwestern.edu
Purpose:
To test the hypotheses that a single internal guide wire coil (i) permits independent and direct depiction of guide wires and catheters and (ii) improves catheter-tracking accuracy and depiction compared to external receiver coils.
Materials And Methods:
Standard 5-6-F angiographic catheters were filled with dilute 4% gadolinium chelate. A single 0.030-inch-diameter internal guide wire coil was placed inside the catheter. True fast imaging with steady-state precession was used to directly visualize the guide wire. Inversion recovery-prepared fast low-angle shot technique was used to track catheters over a thick slice. In phantom experiments, we compared catheter signal-to-noise ratios (SNRs) with the internal coil and a phased-array surface coil with use of the Wilcoxon signed-rank test. Tip-tracking accuracy was assessed with use of linear regression. In pigs (n = 7), catheters and guide wires were independently tracked in real time.
Results:
In phantoms, catheter SNR with the internal coil (12.0) was significantly greater than that with the surface coil (4.0; P =.001). Tip-tracking accuracy was also improved with use of the internal coil (R(2) = 0.94 vs 0.50). In swine vasculature, catheters and guide wires could be directly and independently tracked at 1.7-2.0 frames per second. Catheters were clearly visualized with use of the internal coil, with a typical catheter background contrast-to-noise ratio of 6.6. Catheters were not visible with use of the external coil because of the small catheter size compared to the slice thickness.
Conclusion:
Internal guide wire coils permit independent and direct depiction of guide wires and catheters in vivo for MR imaging-guided endovascular interventions. They also improve catheter tracking accuracy and depiction compared to external coils.