Air leaks during mechanical ventilation as a cause of persistent hypercapnia in neuromuscular disorders
Jesus Gonzalez1, Tarek Sharshar, Nicholas Hart
1Service de Réanimation Médicale, Hôpital Raymond Poincaré, AP-HP, 92380, Garches, France.
Objective:
This study investigated whether air leaks from the upper airway during assisted ventilatory support are associated with persistent hypercapnia (PaCO(2) >45 mmHg) in patients with neuromuscular disorders.
Setting:
A rehabilitation hospital.
Patients:
The study was performed in 95 neuromuscular patients; 52 were tracheostomized with a cuffless tracheostomy tube (invasive ventilation), and 43 received noninvasive ventilation.
Measurements And Results:
The volume of air leaked (VL) and arterial carbon dioxide (PaCO(2)) were routinely measured during mechanical ventilation; PaCO(2) was also measured during spontaneous breathing. VL, expressed as a percentage of tidal volume, was higher in the hypercapnic group (32+/-14%, n=20) than the nonhypercapnic group ( vs. 20+/-14%). PaCO(2) during mechanical ventilation was correlated with both VL and the duration of ventilatory support per day; PaCO(2) during spontaneous breathing was correlated only with the volume of air leaked. In stepwise multiple regression analysis, air leaks contributed to 8% of the variance in PaCO(2) during mechanical ventilation, and daily duration of ventilatory support contributed 5%. In addition, reduction in VL with normalization of PaCO(2) was achieved in five of the noninvasively ventilated patients with persistent hypercapnia by using a chin strap.
Conclusions:
Air leaks during wakefulness are an important cause of persistent hypercapnia in both invasively and noninvasively ventilated neuromuscular patients. However, simple practical measures to reduce the volume of air leaks improve the efficacy of ventilation in these patients.
Related Concept Videos
Acute Respiratory Failure-III
Hyperpnea and Hyperventilation
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)