Related Experiment Video
Updated: Aug 18, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Measurement of functional residual capacity by nitrogen washout during partial ventilatory support
Jörg Zinserling1, Hermann Wrigge, Dirk Varelmann
1Klinik und Poliklinik für Anästhesiologie und Spezielle Intensivmedizin, Rheinische Friedrich-Wilhelms-Universität Bonn, Sigmund-Freud-Strasse 25, 53105 Bonn, Germany. jzinserling@uni-bonn.de
Objective:
Evaluation of an open circuit multiple breath nitrogen washout (MBNW) technique for measurement of functional residual capacity (FRC) during partial ventilatory support using corrections for gas viscosity, sampling delay time, and re-inspired nitrogen.
Design:
Measurements in a lung model with known reference volume simulating spontaneous breathing and duplicate measurements in patients breathing spontaneously with partial ventilatory support. SETTING. Experimental laboratory and intensive care units of a university hospital.
Patients:
Eighteen patients with acute respiratory failure.
Interventions:
Change of FiO(2) from baseline to 1.0.
Measurements And Main Results:
FRC was measured by MBNW during spontaneous breathing with continuous positive airway pressure, pressure support ventilation, proportional assist ventilation, automatic tube compensation, and airway pressure release ventilation. In the lung model, repeated measurements at three volumes were done with all partial ventilatory support modalities, and baseline FiO(2 )was varied with one mode and FRC. The mean of differences between MBNW (FRC(MBNW)) and reference was 28 ml (1.6%), and the 2.SD-interval was 84 ml (4.9%) for all modes. Measurements up to a baseline FiO(2) of 0.8 showed differences of 5 ml (-0.3%) and the 2.SD-interval of 38 ml (2.2%) between reference and FRC(MBNW). In 18 patients, 66 duplicate measurements revealed a mean difference of 30 ml (0.9%) with a coefficient of repeatability of 358 ml (13%) independent of ventilatory mode and chronological order.
Conclusion:
This study suggests that, using corrections for gas viscosity, sampling delay time, and re-inspired nitrogen, FRC can be determined with good repeatability in patients and good accuracy in a lung model during partial ventilatory support.
Related Concept Videos
Lung Capacity
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Respiratory Volumes and Capacities
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...