PET imaging with yttrium-86: comparison of phantom measurements acquired with different PET scanners before and after
H G Buchholz1, H Herzog, G J Förster
1Department of Nuclear Medicine, Johannes Gutenberg-University Hospital Mainz, Mainz, Germany. buchholz@nuklear.klinik.uni-mainz.de
Abstract:
Quantitative imaging with the positron emitter (86)Y is the method of choice to determine the uptake and dosimetry of (90)Y-labelled radiopharmaceuticals. To examine the quantitative accuracy of positron emission tomography findings with (86)Y, this non-pure positron emitter was evaluated in a cylindrical phantom with rods of Teflon, water and air and measured with three different scanners: ECAT EXACT (2D/3D), ECAT HR+ (2D/3D) and PC4096+ (2D). After standard reconstruction, (86)Y radioactivity measured with the ECAT EXACT and related to the true radioactivity varied between 0.84 and 0.99 in 2D and between 0.93 and 1.20 in 3D from the first to the last acquisition (eight half-life times later). The water and Teflon rods exhibited considerable amounts of reconstructed radioactivity-21% in 2D and 67% in 3D for water and 65% and 147%, respectively, for Teflon-compared with the actual (86)Y radioactivity of the phantom. For the ECAT HR+ similar results were obtained in 3D, but there were even greater overestimations in 2D. Measurements with the PC4096+ showed rather small errors, with 10% for water and 20% for Teflon. To correct for the background of gamma-coincidences, sinograms were analysed and an experimental percentage of the background was subtracted from the sinograms. In order to minimise the errors in reconstructed radioactivity, the subtraction value had to be different for the individual scanners and modes. Our results demonstrate that (90)Y/(86)Y-based dosimetry for bone and red marrow must be regarded with caution if it is derived from regions of interest over the bone, the density of which is similar to that of Teflon. To obtain more reliable estimates, an appropriate background correction must be applied and tailored individually with respect to the scanner and acquisition mode.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
![Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F62460.jpg&w=3840&q=50)