Conditioning data for calculation of the modulation transfer function.

Andrew D A Maidment1, Michael Albert

  • 1Department of Radiology, Thomas Jefferson University, Suite 3390 Gibbon Building, 111 South 11th Street, Philadelphia, Pennsylvania 19107-5563, USA. Andrew.Maidment@mail.tju.edu

Medical Physics
|February 28, 2003
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Correction: Scalable flow synthesis of ultrasmall inorganic nanoparticles for biomedical applications via a confined impinging jet mixer.

Scientific reports·2026
Same author

Lesion detectability and masking disparity assessment in breast tomosynthesis across diverse populations using in-silico imaging trials.

IEEE transactions on medical imaging·2026
Same author

Cost-Effectiveness of Artificial Intelligence in Breast Cancer Screening: An Ethical Perspective on a Complex Issue.

Value in health : the journal of the International Society for Pharmacoeconomics and Outcomes Research·2026
Same author

Quantitative evaluation of spatially-variant deformations recovered by deep learning on clinical-like breast lesions.

IEEE transactions on radiation and plasma medical sciences·2026
Same author

Scalable flow synthesis of ultrasmall inorganic nanoparticles for biomedical applications via a confined impinging jet mixer.

Scientific reports·2026
Same author

Scalable flow synthesis of ultrasmall inorganic nanoparticles for biomedical applications <i>via</i> a confined impinging jet mixer.

bioRxiv : the preprint server for biology·2025
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

This study introduces a new data conditioning method for measuring the modulation transfer function (MTF). By ensuring the edge spread function (ESF) is monotonic, this technique improves MTF measurement accuracy for digital detectors.

Area of Science:

  • Medical physics
  • Image analysis
  • Digital imaging

Background:

  • Accurate measurement of the modulation transfer function (MTF) is crucial for evaluating imaging system performance.
  • Traditional MTF measurement methods can be sensitive to noise and artifacts in the data.
  • The edge spread function (ESF) is a key component in calculating the MTF.

Purpose of the Study:

  • To present a novel data conditioning method for MTF measurement.
  • To improve the accuracy and reliability of MTF calculations.
  • To demonstrate the applicability of the method in clinical settings.

Main Methods:

  • A data conditioning technique is proposed based on the monotonic constraint of the edge spread function (ESF).
  • Simulated data with known true MTF values were used to validate the method.

Related Experiment Videos

  • The technique was applied to measure the MTF of a digital detector in clinical use.
  • Main Results:

    • The proposed method demonstrated advantages in MTF measurement accuracy when the ESF constraint was applicable.
    • Simulated examples showed reliable MTF recovery using the conditioning technique.
    • Successful application in measuring the MTF of a clinical digital detector was achieved.

    Conclusions:

    • The monotonic ESF constraint provides an effective data conditioning approach for MTF measurement.
    • This method enhances the accuracy of MTF calculations, particularly for digital detectors.
    • The technique shows promise for routine MTF assessment in clinical imaging applications.