Bilirubin quantitation with lipemic plasma.
Clinical Biochemistry
|April 1, 1976
Summary
Lipemia interferes with bilirubin (BR) measurements, making the diazo method unreliable. A correction factor using the A-O bilirubinometer can improve accuracy for lipemic plasma BR levels.
Area of Science:
- Clinical Chemistry
- Biochemistry
- Laboratory Medicine
Background:
- Accurate measurement of bilirubin (BR) is crucial for diagnosing and monitoring various medical conditions.
- Lipemia, a condition of abnormal lipid levels in the blood, can interfere with spectrophotometric assays, including BR estimation.
- Existing methods for BR determination may yield inaccurate results in the presence of lipemia.
Purpose of the Study:
- To evaluate the reliability of bilirubin estimation in lipemic plasma.
- To assess the performance of the A-O bilirubinometer in the presence of lipemia.
- To develop a method for correcting bilirubin measurements affected by lipemia.
Main Methods:
- Bilirubin (BR) levels were measured in lipemic plasma using the diazo method and the A-O bilirubinometer.
- Theoretical BR values were used as a reference for comparison.
- Regression analysis was performed to correlate the percent error in BR estimation with triglyceride (TG) concentrations.
Main Results:
- The diazo method for BR estimation showed variable and unreliable results in the presence of lipemia.
- The A-O bilirubinometer consistently underestimated BR levels in lipemic plasma compared to theoretical values.
- A linear relationship was observed between the percent error in BR estimation and TG concentrations.
Conclusions:
- The diazo method is not suitable for accurate BR determination in lipemic plasma.
- The A-O bilirubinometer provides lower BR values in lipemic samples.
- A correction factor derived from regression analysis can be applied to adjust BR measurements in lipemic plasma (5-25 mg/dl) when the degree of lipemia is known.