Assessment of image quality of a standard and two dose-reducing protocols in paediatric pelvic CT
John Ratcliffe1, Cheryl E Swanson, Niru Hafiz
1Department of Radiology, The Royal Children's Hospital, Brisbane, Queensland, Australia. ratcliffejohn@hotmail.com
Insights
Reduced radiation dose in paediatric pelvic CT scans using 75 mAs maintains diagnostic quality. Lower mAs settings can be safely used for paediatric pelvic imaging, reducing potential risks.
Area of Science:
- Radiology
- Medical Imaging
- Paediatric Imaging
Background:
- Radiation dose from paediatric pelvic CT scans raises concerns about neoplasm and teratogenic risks.
- Optimizing radiation exposure in paediatric patients is crucial for long-term health outcomes.
Purpose of the Study:
- To evaluate the diagnostic image quality of paediatric pelvic CT scans using reduced mAs or increased pitch.
- To compare image quality against standard CT imaging protocols.
Main Methods:
- A prospective study involved 105 paediatric patients undergoing pelvic CT scans.
- Three protocols were used: standard (200 mA), increased pitch, and reduced mAs (100 mA).
- Radiologists assessed image quality blinded to the protocol used.
Main Results:
- No significant difference in image quality was observed between standard settings and half the standard mAs (75 mAs).
- Increased pitch factor led to a higher rate of suboptimal image quality.
- 120 kVp with 75 mAs yielded diagnostic quality equivalent to 150 mAs.
Conclusions:
- Reduced mAs settings (75 mAs at 120 kVp) provide diagnostic quality comparable to higher settings (150 mAs) for paediatric pelvic CT.
- Lower radiation dose protocols are effective for paediatric pelvic imaging and should be considered standard.
- Optimized CT protocols can minimize radiation risks in paediatric patients.
Background:
Concerns exist regarding the effect of radiation dose from paediatric pelvic CT scans and the potential later risk of radiation-induced neoplasm and teratogenic outcomes in these patients.
Objective:
To assess the diagnostic quality of CT images of the paediatric pelvis using either reduced mAs or increased pitch compared with standard settings.
Materials And Methods:
A prospective study of pelvic CT scans of 105 paediatric patients was performed using one of three protocols: (1) 31 at a standard protocol of 200 mA with rotation time of 0.75 s at 120 kVp and a pitch factor approximating 1.4; (2) 31 at increased pitch factor approaching 2 and 200 mA; and (3) 43 at a reduced setting of 100 mA and a pitch factor of 1.4. All other settings remained the same in all three groups. Image quality was assessed by radiologists blinded to the protocol used in each scan.
Results:
No significant difference was found between the quality of images acquired at standard settings and those acquired at half the standard mAs. The use of increased pitch factor resulted in a higher proportion of poor images.
Conclusions:
Images acquired at 120 kVp using 75 mAs are equivalent in diagnostic quality to those acquired at 150 mAs. Reduced settings can provide useful imaging of the paediatric pelvis and should be considered as a standard protocol in these situations.
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies V: Intravenous Urography and Retrograde Pyelography